RINSE2 STANDARD
VALUE (type)
PF10 RINSE2 SELECTOR: value 0.1
Rinse 2 selector (Main parameter for selecting or deselecting the whole
rinse 2.)
0 (P)
CT05 RINSE2 FILLTYPE: value 1, 2, 3, 4, 5 (CW, HW, CW+HW, booster, HCW)
Rinse 2 water type selector (parameter for selecting water to be used in
rinse 2
1=cold water, 2=hot water, 3=mixture of cold and hot water, 4=filling from
booster tank, 5=high quality water).
2 (P)
LI10 RINSE2 TEMP SP: value 0-90°C
Rinse 2 temperature set point.
0°C (P)
TI07 RINSE2 TIME: value 0-60 min
Rinse 2 time at selected rinse 2 temperature set point.
1 min (P)
CT14 RINSE2 LOOPS: value 1-3
Rinse 2 number of desired loops 1-3.
1 (P)
PF13 RINSE2 GUARANTEE: value 0.1
Rinse temperature guarantee selector 2. (0=rinse time starts directly
whatever temperature achieved, 1=rinse time starts when selected rinse
temperature is reached)
1 (P)