General recommendations for non Getinge detergents
Type Recommended pump hose
Alkaline detergents 503611700, Norprene
Acidic detergents (dishwasher detergent, fabric
softener, neutral)
503611703, Silicone
Enzymes 503611700, Norprene
Lubricants 503611702, Tygon
Variations may occur as the chemical composition of different detergents on the market can vary
considerably. For example, Silicone 503611703 is recommended for acidic agents based on citric acid,
while Fluran 503611704 is recommended for acidic agents based on phosphoric acid. Contact the
current detergent supplier if uncertain about which hose to select.
Setting detergent and rinse-aid quantities
This must only be done by authorized personnel.
The machine is connected to the electricity supply and some components are live.
Dispense detergent or rinse-aid into a measuring beaker. Check that the suction hose and
pump are full before inspection.
Push the suction hose down into the beaker and measure the consumption during an entire
cleaning program. Lift up the hose when reading the volume. Adjust if necessary, and repeat
the measurement until the amount of detergent conforms to the manufacturer’s
recommendations. See "Calibrating dosing monitor – set volume (2.1.8.2.3.2)" on page 40.
000545
Page 138 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals