Set the limit
1. Press "E" on the transmitter.
2. Enter code 22 with the "+" and "-" buttons.
3. Press "E". "Setup 1" appears on the display.
4. Choose "relay" with the "+" and "-" buttons.
5. Hold down "E" until "A. Thresh" appears on the display.
6. Enter 9999 with the "+" and "-" buttons.
7. Press "E".
8. Press "+" and "-" simultaneously to log out.
Adjusting analog input
1. Press "E" on the Conductivity transmitter.
2. Enter code 22 with the "+" and "-" buttons.
3. Press "E". "Setup 1" appears on the display.
4. Choose "Output" with the "+" and "-" buttons.
5. Press "E". Release the button when "Sel. type" appears.
6. Choose "SIM" with the "+" and "-" buttons.
7. Press "E".
8. Select 0...20 mA with the "+" and "-" buttons.
9. Enter 4 mA.
See section Automatic calibration (2.1.6.2.2), Page 48
Changing the display.
Press "+" to switch the display between:
• Non-compensated value (µS/cm)
• Temperature (°C)
• Temperature (°F)
• No value
Page 131 of 166