FINAL RINSE / DISINFECTION STANDARD
VALUE (type)
PF04 FRINSE SELECTOR: value 0.1
Final rinse selector (Main parameter for selecting or deselecting the whole
final rinse.)
0 (P)
CT04 FRINSE FILLTYPE: value 1, 2, 3, 4, 5 (CW, HW, CW+HW, booster, HCW)
Final rinse water type selector (parameter for selecting type of water to be
used in final rinse.
1=cold water, 2=hot water, 3=mixture of cold and hot water, 4=filling from
booster tank, 5=high quality water).
2 (P)
LI03 FRINSE TEMP SP-value 20-90°C
Final rinse temperature set point.
90°C (P)
TI04 FRINSE TEMP TIME: value 0-60 min
Final rinse time at selected final rinse temperature set point.
1 min (P)
PF14 P RINSE GUARANTEE: value 0.1
Selector for final rinse temperature guarantee. (0=final rinse time starts
directly whatever temperature achieved, 1= rinse time starts when set
rinse temperature is achieved)
1 (P)
LI09 FRINSE DOS TEMP: value 0-85°C
Final rinse dose temperature set point.
85°C (P)
CT08 FRINSE DOS PUMP: value 0, 1, 2, 3, 4
Final rinse dose pump type selector (Parameter for selecting type of pump
to be used in the final rinse dose,
0=no dose, 1= dose pump, 1 2=dose pump 2 3=dose pump 3 4=dose
pump 4).
2 (P)
PF39 FRINSE CC: value: 0.1
Conductivity control of the final rinse (Main parameter for selecting or
deselecting the conductivity control of the final rinse.)
0 (P)
LI50 FRINSE COND SP: value 0-500 µS/cm
Conductivity final rinse set point.
Valid if PF39 is 1 only.
250 µS/cm (P)
PF30 DISINF: value 0.1
Disinfection selector (Main parameter for selecting or deselecting the
disinfection.)
Possibility to deselect the nomenclature disinfection.
1 (P)
PF34 DISINF Ao PRINT: value 0.1
Disinfection Ao printed (Selector if Ao value printed.)
1 (P)
PF35 DISINF Ao CONTR: value 0.1
Disinfection Ao controlled (Selector if disinfection is controlled by Ao
instead of temperature and time.)
1 (P)
LI51 DISINF Ao SP: value 0-6000 sec
Disinfection Ao set point.
Valid if PF35 is 1 only.
600 sec. (P)