WASH2, NEUTRALIZING, CHEM DISINFEC. STANDARD
VALUE (type)
PF02 FWASH WASH2 SEL: value 0.1
Wash 2 selector (Main parameter for selecting or deselecting the whole
wash 2.)
One and only one of the three selectors Wash 2, Neutralizing, Chemical
disinfection can be selected at the same time.
0 (P)
PF08 FWASH NEUTR SEL: value 0.1
Neutralizing (goods) selector (Main parameter for selecting or deselecting
the whole neutralizing.)
One and only one of the three selectors Wash 2, Neutralizing, Chemical
disinfection can be selected at the same time.
0 (P)
PF09 FWASH CHEM SEL: value 0.1
Chemical disinfection selector (Main parameter for selecting or deselecting
the whole chemical disinfection.)
One and only one of the three selectors Wash 2, Neutralizing, Chemical
disinfection can be selected at the same time.
0 (P)
CT02 FWASH FILLTYPE: value 1, 2, 3, 4, 5 (CW, HW, CW+HW, booster, HCW)
Water type for final wash 1 (parameter for selecting type of water to be
used during final was.
1=cold water, 2=hot water, 3=mixed cold and hot value, 4=Filling from
booster tank, 5= High quality water).
3 (P)
LI02 F WASH TEMP SP-value 20-90°C
Final Wash temperature set point.
60°C (P)
TI02 FWASH TIME: value 0-60 min
Final Wash time at selected wash 2 temperature set point
2 min (P)
LI08 FWASH DOS TEMP: value 0-60°C
Final Wash detergent dosing temperature set point.
45°C (P)
CT07 FWASH DOS PUMP: value 0, 1, 2, 3, 4
Final Wash detergent dosing pump selector (o=no pump, 1=pump 1,
2=pump 2, 3=pump 3, 4=pump 4)
2 (P)
PF21 FWASH TANK2: value 0.1
Switch for saving tank for final wash. (0=no saving tank used for filling and
draining final wash,
1=saving tank 2 used when filling and emptying final wash) (saving tank 1
always used for final wash 1, saving tank 2 always for final wash)
1 (P)
PF28 FWASH DOS TYPE: value 0-1 (normal amount based, Conductivity
monitor)
Switch for type of detergent dosing for final wash. (0=detergent dose
based on normal amount, 1=Conductivity controlled, detergent dosing
based on set conductivity of the water and detergent mixture)
0 (P)
Page 72 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals