Automatic calibration (2.1.6.2.2)
49,1 C
WASH1 TEMP SP
P1 HC SHORT 00:12:24
01 STANDBY 00:02:22
A34 I/O FAULT
00:03:00
WASH1 TIME
128,1 C
DRYER TEMP
CANCEL
AUTOMATIC CALIBRATION
SENSOR CALIBRATE
00 CHAMBER TEMP YES
01 DRYER TEMP NO
02 DRAIN TEMP NO
03 BOOSTER TEMP NO
05 DISCHARGE PRESS NO
06 CONDUCTIVITY FR NO
000725
2
1
1. CALIBRATE YES/NO
2. CANCEL
Default value is NO. When any sensor is set to YES and this is confirmed with ENTER, the NEXT button
appears in the bottom right corner. NEXT opens the AUTOMATIC CALIBRATION HIGH/LOW (2.1.8.2.2.2)
screen.
If a sensor is set to YES, the other sensors change color and cannot be selected.
Page 48 of 166