Test digital output (2.1.6.8.2.4)
49,1 C
WASH1 TEMP SP
P1 HC SHORT 00:12:24
01 STANDBY 00:02:22
A34 I/O FAULT
00:03:00
WASH1 TIME
128,1 C
DRYER TEMP
CANCEL HOME OK
000762
TEST DIGITAL OUTPUT
11
01
02
03
04
05
CIRC PUMP
DRAIN PUMP
DET PUMP 1
DET PUMP 2
FAN DRYER
DOOR SS THR
OUTPUT MODE ACT. VAL
AUT
AUT
AUT
AUT
AUT
AUT
SET
0
0
0
1
1
1
0
0
0
0
0
0
MAN
MOD
3
2
1
1. MAN MOD
2. CANCEL
3. A34 I/O FAULT
MAN MODE
When activating MAN MODE three functions appear:
CANCEL:
General function
AUTO/MAN: Toggle auto-man
I/O Toggle the output between 1 (Active) and 0 (Inactive)
PULSE Activates the output for 5 sec after 10 sec delay
ACTIVATE: Activates all manual set outputs
If leaving this menu in MAN MODE, a confirmation has to be done. ”CONFIRM MAN MODE SET” Yes or
No.
Page 95 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents