DESCALING STANDARD
VALUE (type)
PF05 DESCALE SELECTOR value 0.1
Descaling selector (main parameter for activating or deactivating the
descaling).
To be used with emptying/drying only. In order to create a descaling
cycle. Cannot be combined with other phases.
Descaling includes saving tank 1 if DS24 is 1. (saving tank 1 is installed)
0 (P)
TI17 DESCALING TIME: value 0-60 min
Descaling time.
10 min (P)
AFTER TREATMENT STANDARD
VALUE (type)
PF11 ATREAT SEL: value 0.1
After treatment selector (Main parameter for selecting or deselecting the
after treatment.)
0 (P)
PF37 ATREAT AWRFS: value 0.1
After treatment with automatic water rack flush system (main parameter
for activating or deactivating after treatment with automatic water rack
flush system).
1 (P)
PF38 ATREAT DFR: value 0.1
After treatment direct final rinse (Main parameter for selecting or
deselecting the after treatment direct final rinse.)
1 (P)
TI20 ATREAT AWRFSTIME: value 0-2 min
Time for after treatment with automatic water rack flush system.
Valid if PF37 is 1 only.
2 min (P)
TI12 ATREAT DFR TIME: value 0-2 min
After treatment direct final rinse time.
Valid if PF38 is 1 only.
2 min (P)
Page 76 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals