DRAIN/DRYING STANDARD
VALUE (type)
PF07 DRYING SELECTOR: value 0.1
Drying selector (Main parameter for selecting or deselecting the drying.)
Drying cannot be deselected; only the type of draining can be selected
(See PF32).
1 (P)
TI05 DRY TIME: value 0-30 min
Drying time.
5 min (P)
LI04 DRY TEMP SP: value 20-120°C
Set drying time temperature
110°C (P)
LI06
DRY MAX CHAMBER: value 0-90°C
Maximum drying temperature in the chamber. Normally reduced when
sterilizing items that cannot handle temperatures above 90°C.
90°C (P)
PF33 DRY TILT: value 0.1
Dry tilt (Parameter for selecting or deselecting the tilt.)
1 (P)
PF32 FINAL DRAIN TYPE: value 0.1 (normal, partly)
Final drain type (parameter for selecting normal or partial drain for the final
draining.)
If normal is selected, all the contents of the wash drain is emptied.
If partial is selected the chamber drains until CHAMBER LEVEL H has lost
its signal. (about 60 l left in the sump) Note! This is normally very hot water
that will be reused in the next cycle.
0 (P)
PF23 DRAIN TANK1: value 0.1
Drain tank 1 (Parameter for selecting or deselecting draining of tank 1.)
Main purpose for this parameter is a possibility to create drain cycles.
0 (P)
PF24 DRAIN TANK2: value 0.1
Drain tank 2 (Parameter for selecting or deselecting draining of tank 2.)
Main purpose for this parameter is a possibility to create drain cycles.
0 (P)
Page 77 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>