In an emergency
• Press the emergency stop.
• Switch off the main power switch.
• Close shutoff valves in the water and any steam supply lines.
Make sure that the assembly and installation is carried out:
• by qualified personnel.
• in accordance with current local regulations and rules.
To avoid the risk of back injury, this equipment should be assembled and installed
by at least two people.
When the work is complete, check that
• all parts have been installed according to the installation manual.
• all screws have been properly tightened
• there are no sharp edges on any parts that may come into contact with people
• all hoses, pipes and connections are intact and free from defects
• all the functions of the machine are working properly. Adjust if necessary.
• make sure that no electrical wires or hoses are touching any hot surfaces
• check electrical connections both inside the electrical cabinet and to elements
Carry out an Installation Qualification, an Operating Qualification, and a Performance Qualification
according to ISO 15883 before putting the machine into service.
Incorrect use may result in damage to objects and personal injury.
Product liability
Any modification or incorrect use of the equipment without the approval of Getinge Disinfection AB
negates Getinge Disinfection AB’s product liability.