PEL-3000(H) User Manual
150
Power Source
Config Error
FSEQRS232
03/Oct/2018
OCP Test
0.00
V
0.000
A
0.00
w
Config Error
Config Error indicates that upon
starting the power supply voltage is
less than the Trig V voltage.
Set Curr: Fetch Volt:
Config Error indicates that the power supply
voltage is less than the Trig V voltage setting after
the test has started. This can indicate that the
power supply output is not on or that the power
supply output or Trig V is incorrectly configured.
In addition to the OCP settings as described above,
the Trig voltage settings must also be set according to
the output characteristics of the DUT.