360077-1/O 112
Horizon Series™ - Laboratory, Pharmacy, and International Blood Bank Models
24.3.1 High Temperature Alarm
► SpeciesthetemperatureatwhichtheHighTemperatureAlarmactivates.
► Default setpoint is 5.5 °C.
► Setpoint can be changed from -40.0 °C to +25.0 °C.
Change the setpoint:
1 Hold Up Arrow and Down Arrowforthreeseconds.MONITORlampwillashtoindicateentryinto
program mode.
2 Press SELuntilHIGHTEMPandMONITORlampsash.
3 Hold SET, then press Up Arrow or Down Arrow to change the setpoint.
4 Release SET button.
5 Hold Up Arrow and Down Arrowforthreeseconds.MONITORlampstopsashingtoindicateexit
from program mode. New settings are saved.
24.3.2 Low Temperature Alarm
► SpeciesthetemperatureatwhichtheLowTemperatureAlarmactivates.
► Default setpoint for HHB (blood bank) models is 1.5 °C.
► Default setpoint for HLR and HPR (laboratory and pharmacy) models is 2.0 °C.
► Setpoint can be changed from -40.0 °C to +25.0 °C.
Change the setpoint:
1 Hold Up Arrow and Down Arrowforthreeseconds.MONITORlampwillashtoindicateentryinto
program mode.
2 Press SELuntilLOWTEMPandMONITORlampsash.
3 Hold SET, then press Up Arrow or Down Arrow to change the setpoint.
4 Release SET button.
5 Hold Up Arrow and Down Arrowforthreeseconds.MONITORlampstopsashingtoindicateexit
from program mode. New settings are saved.
24.4 Temperature Calibration Setpoints
View setpoints:
1 Hold Up Arrow and Down Arrowforthreeseconds.MONITORlampwillashtoindicateentryinto
program mode.
2 Press SEL until desired setting appears.
3 Observe the setting.
4 To view another setting, press SEL again (optional).
5 Hold Up Arrow and Down Arrowforthreeseconds.MONITORlampstopsashingtoindicateexit
from program mode.
Flashing Lamp Selected Setting
HIGH TEMP and MONITOR High temp alarm setpoint
LOW TEMP and MONITOR Low temp alarm setpoint
MONITOR only Monitor offset
CONTROL only Control sensor offset
CONTROL only Control hysteresis