Test the alarm:
1 Ensure the monitoring system backup batteries are installed.
2 Remove one battery from the monitoring system backup battery holder.
a If the no battery alarm activates, no further action is needed. Reinstall the battery.
b If the no battery alarm does not activate, contact Helmer Technical Service.
NOTICE Wheninstallingreplacementbatteries,useonlybatterieswhichmeetthespecications
outlined in chapter 10.7 (Supplies).
8.10 Additional System Settings
8.10.1 Screen Contrast
The screen contrast can be changed for easier viewing.
NOTE ► During an AC power failure, the screen backlight is not illuminated to conserve
backup battery power.
► During an AC power failure, the screen contrast cannot be changed.
Change screen contrast:
1 On the HOME screen, press the third button from the left to make the text appear lighter.
2 On the HOME screen, press the second button from the left to make the text appear darker.
8.10.2 Date and Time
The Date Format setting controls the order in which the month (mm) and day (dd) are displayed.
► Month is a 2-digit number (01-12)
► Day is a 2-digit number (01-31)
► Default date format is mm/dd/yyyy
The Clock Mode setting controls whether the time is displayed in a 12-hour or 24-hour format.
► Whenusingthe12-hourformat,AMorPMmustbespecied
► Default setting is 12-hour
Change date and time settings:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Set Date & Time. Press the SELECT button.
5 Press the UP or DOWN buttons to select the date and time settings to change.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.10.3 Display Language
The i.Center monitoring system stores two languages. English is the default language. If a different
languageisdesired,itmustbeloadedfromtheashmemorycard.Ifaashmemorycardisnotincluded
with the refrigerator, the languages may have been loaded to the i.Center prior to shipment.
Toobtainaashmemorycardtoloadanalternatelanguage,contactHelmerTechnicalService.
NOTE Each time the refrigerator is powered on, the i.Center display language must be selected.