360077-1/O 26
i.Series® - All Models
5 Press the DOWN button to highlight Store Calibration.
a To save the new value, press the ENTER button. The “Calibration Memorized” message
appears. New settings are saved.
b To discard the new value, press the BACK button or HOME button to exit. New settings are not
saved.
6 Remove thermometer and probe from bottle.
7 Replace the probe in probe bottle.
8 Replacebottlecap,ensuringatightt.
9 Place the probe in bottle, immersing at least 2” (50 mm).
NOTE ► The current temperature displayed by the monitoring system may change so that it
no longer matches the new probe calibration value. This is normal.
► If a new probe value is entered but not saved, the new value will appear when the
calibration setting for the probe is viewed. This is normal.
8.6 Factory Default Settings
Settings listed below may be simultaneously returned to factory default values.
NOTE The factory default settings may not be the same as the settings that were factory-
calibrated before the refrigerator was shipped.
Setting Restored Value
High Alarm Setpoint 5.5 °C
Low Alarm Setpoint
(1)
1.5 °C
Condenser Alarm Setpoint 50.0 °C
Door Ajar Timeout 3 minutes
Power Failure Timeout 3 minutes
Chart Paper Timer 6.5 days
(1) This includes laboratory (iLR) and pharmacy (iPR) models that were originally set at 2.0 °C.
NOTE Unless your organization requires the Low Alarm Setpoint to be at the factory default
level of 1.5 °C, it will be necessary to increase the setpoint to 2.0 °C after restoring
factory defaults.
8.7 Restore Factory Default Settings
Restore settings:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Factory Default Settings. Press the SELECT button.
5 Do one of the following:
► Press the ENTER button. Factory default settings are restored.
► Press the BACK button. Factory default settings are not restored.