360077-1/O 29
i.Series® - All Models
3 Enter the password when prompted.
4 Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5 Press the DOWN button to highlight Door Ajar Timeout.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.7 Power Failure Alarm
ThePowerFailureTimeoutspecieslongesttimetherefrigeratorcanbewithoutACpowerbeforethe
alarm activates. If the time elapsed since the last power failure is greater than or equal to this value, the
alarm activates.
► Default delay setting is three minutes
► Setting can be changed from 0 minutes to 60 minutes
Change the alarm delay:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5 Press the DOWN button to highlight Power Failure Timeout.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.8 Chart Paper Alarm
The default setting for the chart paper timer is Enabled. One sheet of chart paper records temperatures
continuously for seven days. The timer activates an alarm 6.5 days from when the timer is reset. The
timer period cannot be changed.
NOTE ► Available options are Enabled, Disabled, and Reset.
► Enabling the timer also resets the timer.
Change the setting:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight System Options. Press the SELECT button.
5 Press the DOWN button to highlight Chart Paper Timer.
6 Press the INC or DEC buttons to select Enabled, Disabled, or Reset.
7 Do one of the following:
► If Enabled or Disabled is selected, press the BACK button to return to the System Options
screen, or press the HOME button to exit. The new setting is saved.
► If Reset is selected:
a Press the DOWN button.
b Press the PAPER-CHANGED button. The System Options screen appears with the Chart
Paper Timer set to Enabled.
8 Press the BACK button to return to the System Options screen, or press the HOME button to exit.
The new setting is saved.