360077-1/O 72
Horizon Series™ - Blood Bank Models
16.3 ChangeCongurationPassword
The default password is 1234. A new password must use four digits, ranging from 1 to 5.
Change the password:
1 On the HOME screen, press the MAIN button.
2 Press the DOWNbuttontoselectEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to select Change Password. Press the SELECT button.
5 Enter the new password, then re-enter the new password when prompted.
► If password entries match, the “update” message is displayed.
► If password entries do not match, the “incorrect match” message is displayed. Repeat the
procedure to change the password.
16.4 Calibrate Chamber Temperature Probe
Verify the temperature probe is reading chamber temperature correctly by comparing the chamber probe
reading to temperature read by an independent thermometer. If the chamber temperature probe is not
reading correctly, change the value displayed on the temperature monitoring system.
NOTE If the variance is within acceptable limits for your organization, changing probe settings is
optional.
► Default setting for chamber temperature is 4.0 °C
► Value is factory-preset
Obtain:
► Independent thermometer, calibrated and traceable per national standards
Measure the chamber temperature:
1 Remove the probe from the probe bottle.
2 Unscrew the cap from the bottle.
3 Insert the thermometer and temperature probe in the bottle. The probe and thermometer should be
immersed at least 2” (50 mm).
4 Close the door and allow the chamber temperature to stabilize for 10 minutes.
5 Observe and note the thermometer temperature.
EXAMPLE ► Measured temperature (at the probe bottle) is 4.0 °C
► Displayed temperature is 4.5 °C
► Change displayed temperature to 4.0 °C
Enter the new calibration value:
1 On the HOME screen, press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Temperature Calibration. Press the SELECT button.
a The Upper Probe option is displayed and the Temperature option is highlighted.
b Press the INC or DEC buttons to change the displayed value to match the value measured by the
independent thermometer.