360077-1/O 28
i.Series® - All Models
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.4 Low Chamber Temperature Alarm
TheLowAlarmsetpointspeciesthetemperatureatwhichtheLowTemperatureAlarmactivates.Ifthe
temperature detected by the chamber probe is less than or equal to this value, the alarm activates.
► Default setpoint is 1.5 °C (iB models)
► Default setpoint is 2.0 °C (iLR models)
► Setpoint can be changed from -40 °C to +40 °C
Change the setpoint:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5 Press the DOWN button to highlight Low Alarm Setpoint.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.5 Condenser Temperature Alarm
TheCondenserAlarmsetpointspeciesthetemperatureatwhichtheCondenserTemperatureAlarm
activates. If the temperature of the condenser discharge line is greater than or equal to this value, the
alarm activates.
► Default setpoint is 50 °C
► Setpoint can be changed from -40 °C to +80 °C
NOTICE Condenser Temperature Alarm should not be changed unless directed by Helmer
Technical Service.
Change the setpoint:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5 Press the DOWN button to highlight Cond. Alarm Setpoint.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.6 Door Ajar Alarm
TheDoorAjarTimeoutspecieslongesttimetherefrigeratordoorcanbeopenbeforethealarmactivates.
If the time elapsed since the last door opening is greater than or equal to this value, the alarm activates.
► Default delay setting is three minutes
► Setting can be changed from 0 minutes to 60 minutes
Change the alarm delay:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.