360077-1/O 78
Horizon Series™ - Blood Bank Models
16.8.2 Power Failure Alarm Test
NOTE ► During a power failure, the power failure alarm activates and the batteries provide
power to the monitoring system.
► If AC power fails, the backup batteries will allow for continued temperature display.
► If the backup batteries fail, the temperature is not displayed.
► When power is restored, the monitoring system resumes data display.
1 ConrmtherefrigeratorisconnectedtoACpower.
2 Ensure the monitoring system backup batteries are installed.
3 Change Power Failure Timeout setting to 0 minutes.
a Press the MAIN button.
b Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
c Enter the password when prompted.
d Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
e Press the DOWN button to highlight Power Failure Timeout.
f Press the DEC button to change the setting to 0.
4 Disconnect the refrigerator from AC power. Power failure alarm will activate immediately.
5 Reconnect the refrigerator to AC power. Power failure alarm will clear and audible alarm will cease.
6 Change the Power Failure Timeout setting to the original setting.
16.8.3 Door Ajar Alarm Test
1 Change Door Ajar Timeout setting to 0 minutes:
a Press the MAIN button.
b Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
c Enter the password when prompted.
d Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
e Press the DOWN button to highlight Door Ajar Timeout.
f Press the DEC button to change the setting to 0.
g Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to
exit. New settings are saved.
2 Open the door. Door ajar alarm will activate immediately.
3 Close the door. Door ajar alarm will clear and audible alarm will cease.
4 Change the Door Ajar Timeout setting to the original setting.