360077-1/O 27
i.Series® - All Models
8.8 Alarm Settings
The following alarm settings may be changed by the operator. The setpoint for temperature alarms may
be changed (where applicable), as well as the time delay between when the alarm condition commences
and when the visual and audible alarms are initiated.
8.8.1 Alarm Volume
The alarm volume can be changed. The Alarm Volume controls volume for all audible alarms.
► Default setting is 10
► Setting can be changed from 1 to 10
► 1 is the quietest setting; 10 is the loudest setting
Change the alarm volume:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight System Options. Press the SELECT button.
5 Press the DOWN button to highlight Alarm Volume.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.2 Alarm Pulse
The alarm pattern can be changed. This is useful if several refrigerators with alarms are collocated, and
distinguishing the source of the alarm quickly is desirable.
► Default setting is Single.
► Setting can be changed between Single, Double, Triple, and Constant.
Change the alarm pulse:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight System Options. Press the SELECT button.
5 Press the DOWN button to highlight Alarm Pulse.
6 Press the INC or DEC buttons to change the setting.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
8.8.3 High Chamber Temperature Alarm
TheHighAlarmsetpointspeciesthetemperatureatwhichtheHighTemperatureAlarmactivates.Ifthe
temperature detected by the chamber probe is greater than or equal to this value, the alarm activates.
► Default setpoint is 5.5 °C
► Setpoint can be changed from -40 °C to +40 °C
Change the setpoint:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button.
5 Press the DOWN button to highlight High Alarm Setpoint.
6 Press the INC or DEC buttons to change the setting.