EasyManua.ls Logo

Helmer iB111 (Version A) - Chart Paper Alarm

Helmer iB111 (Version A)
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
360077-1/O 76
Horizon Series™ - Blood Bank Models
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new settings are saved.
16.7.7 Chart Paper Alarm
The default setting for the chart paper timer is Enabled. One sheet of chart paper records temperatures
continuously for seven days. The timer activates an alarm 6.5 days from when the timer is reset. The
timer period cannot be changed.
NOTE ► Available options are Enabled, Disabled, and Reset.
► Enabling the timer also resets the timer.
Change the setting:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight System Options. Press the SELECT button.
5 Press the DOWN button to highlight Chart Paper Timer.
6 Press the INC or DEC buttons to select Enabled, Disabled, or Reset.
7 Do one of the following:
► If Enabled or Disabled is selected, press the BACK button to return to the System Options
screen, or press the HOME button to exit. The new setting is saved.
► If Reset is selected:
a Press the DOWN button.
b Press the PAPER-CHANGED button. The System Options screen appears with the Chart
Paper Timer set to Enabled.
8 Press the BACK button to return to the System Options screen, or press the HOME button to exit.
The new setting is saved.

Table of Contents

Related product manuals