360077-1/O 120
Horizon Series™ - Laboratory, Pharmacy, and International Blood Bank Models
26 Troubleshooting
CAUTION ► Review all safety instructions prior to troubleshooting. Refer to chapter 2 (Safety).
► Troubleshooting should only be performed by trained refrigeration technicians.
26.1 General Operation Problems
Problem Possible Cause Action
Drawer or basket
does not slide easily.
Drawer slide is faulty. ► Conrmtheslideisoperatingcorrectly.Replaceif
necessary.
Debris in the drawer
slides.
► Pullthedrawerorbasketoutandconrmthe
slides are free of debris. Clean if necessary.
Drawer slides are not
lubricated.
► Using a lightweight oil, lubricate the bearings in the
slides.
Drawer or basket is
misaligned or not
level.
► Conrmbothslidesforthedrawerorbasketare
mounted at the same height.
Door does not open
easily.
Debris in the hinges. ► Conrmthehingesarefreeofdebris.Cleanthe
hinges if necessary.
Door hinges are not
lubricated.
► Using a general-purpose grease, lubricate the
pivots in the hinges.
Hinge cam is faulty. ► Conrmthehingecamisnotdamaged.Replaceif
necessary.
Chamber temperature
meets an alarm
condition, but
the appropriate
temperature alarm is
not active.
Temperature alarm
setpoint was changed.
► Check the current setpoints for the temperature
alarms. Change the setpoints if necessary.