360077-1/O 43
i.Series® - All Models
10.5.3 Interior
Clean painted surfaces with mild detergent. Clean stainless steel surfaces with a general-purpose
laboratory cleaner suitable for stainless steel.
10.5.4 Door Gaskets
Clean with soft cloth and mild soap and water solution.
10.5.5 CleanandRellProbeBottle
NOTE A kit that includes a probe bottle and glycerin is available from Helmer.
Obtain:
► Fresh water-bleach solution (not provided)
► 1:9 ratio of bleach to water
► Bleach is 5% solution of commercial sodium hypochlorite (NaOCl)
► Equivalent oxidizing cleaner/disinfectant approved by your organization may be substituted
► 4 oz. (120 mL) of product simulation solution per bottle
► 10:1 ratio of water to glycerin
Cleanandrellbottle:
1 Remove probe from bottle.
2 Remove bottle from bracket.
3 Clean bottle with water-bleach solution.
4 Fill bottle with 4 oz. (120 mL) of product simulation solution.
5 Cap bottle tightly to minimize evaporation.
6 Place bottle in bracket.
7 Replace probe, immersing at least 2” (50 mm).