4 Press the HOME button to return to the HOME screen.
5 If a temperature alarm sounds, press the MUTE button.
Change the display language:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight System Options. Press the SELECT button.
5 Press the UP or DOWN buttons to select Language. Press the SELECT button.
6 Press the INC or DEC buttons to select the desired language.
7 Press the BACK button to return to the System Options screen, or press the HOME button to exit.
The new settings are saved.
16.9.4 Temperature Units
Available options are Celsius (°C) or Fahrenheit (°F). The default temperature unit is Celsius.
Change temperature units:
1 Press the MAIN button.
2 Press the DOWNbuttontohighlightEditConguration.PresstheSELECT button.
3 Enter the password when prompted.
4 Press the DOWN button to highlight System Options. Press the SELECT button.
5 Press the DOWN button to highlight Temperature Units.
6 Press the INC or DEC buttons to select the desired temperature units.
7 Press the BACKbuttontoreturntotheEditCongurationscreen,orpresstheHOME button to exit.
The new setting is saved.
16.10 Upgrade System Firmware
HelmermayoccasionallyissueupdatesfortheHorizonSeriesmonitoringsystemrmware.Follow
upgradeinstructionsincludedwiththermwareupdate.
16.11 Reset the Horizon Series Monitoring System
1 Remove 1 battery from the monitoring system backup battery holder.
2 Disconnect the refrigerator from AC power.
3 Reconnect the refrigerator to AC power.
4 Reinstall the battery in the monitoring system backup battery holder.
16.12 View Manufacturer and Product Information
1 Press the MAIN button.
2 Press the INC or DEC buttons to select the Product/Company Information option. Press the SELECT
button.
► Manufacturer contact information appears.
► Software version appears.