360179-A/B 47
Troubleshooting
Problem Possible Cause Action
The compressor runs
continuously.
Compartmental
Access Refrigerator
setpoint is set too low
Conrm the setpoint is set within the operating
range and change it if necessary (refer to Section
II, Item 8.3.1).
Temperature control
probe inaccuracy
Verify control o󰀨set value is displaying between -3
to +3.
Value displayed for the control probe is close to
display temperature.
Conrm the probe is providing resistance in the
range of 98 Ω to 110 Ω. Replace the probe if
necessary.
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Compressor solid
state relay is faulty
Conrm the control indicates the compressor
should not be running. Touch i.C³ APPS, Settings
(enter the Settings password), Device Status.
If the compressor should not be running, check the
compressor solid state relay.
If the compressor solid state relay is closed,
replace the relay.
No air ow across
condenser
Clean condenser.
Verify condenser fan motor is operating.
Improper air ow
across evaperator
Verify evaporator fan motors are operating.
i.C³ control board is
faulty
Conrm the control indicates the compressor
should be running. Touch i.C³ APPS, Settings
(enter the Settings password), Device Status.
If the compressor should be running, no further
action is needed.
If the compressor should not be running, check the
control board compressor relay for SSR1. If the
relay is closed, replace the control board.