8.4 Command Reference
────────────────────────────────────────────────────
Synta
data1: standard temperature, data2: temperature coefficient
data1 = -10.0 to 99.9 (numerical data in NRf format)
data2 = -9999 to 9999 (numerical data in NRf format)
Functio
Sets the standard temperature and the temperature coefficient of the
temperature correction function.
Erro
When the temperature conversion is selected at power on, the execution
error is generated.
Exampl
Transmission
:TCSET 23.0, 3650
The standard temperature of the temperature correction function is set to
23.0℃, and the temperature coefficient of that is set to 3650 ppm (copper
wire).
Synta
Returns the settings of the correction temperature and the temperature
coefficient of the temperature correction function, as a numerical value in
NRf format (data).
Erro
When the temperature conversion is selected at power on, the execution
error is generated.
Respons
If headers are ON
If headers are OFF
:TCSET data1,data2
data1,data2
Exampl
Transmission
Response
:TCSET?
:TCSET 23.0,3650
:TCSET?
23.0, 3650
:TCSET
:TCSET?