EasyManua.ls Logo

Hioki 3227 - Page 95

Hioki 3227
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
83
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
*
RST
Functio
n
Resets the 3277 unit, and initializes the settings.
Items which are initialized (the 3277 unit)
Items Setting items
Measurement mode Resistance measurement mode
Resistance measurement range 300mΩ range
Measurement sampling rate SLOW mode
Measurement holding Free-running
Comparator
Buzzer mode
Mode of the external control terminal
Comparison method of the comparator
High limit value/low limit value
All comparator tables are unused
OFF
AUTO
HIGH/LOW
High limit value:00000/
low limit value:00000
Temperature correction function
Standard temperature
Temperature coefficient
Unused
20.0
3930 ppm
Δt function
Initial resistance
Initial temperature
Constant
Unused
0.00 mΩ
0.0
235.0
Use power supply frequency 50 Hz
Items which are initialized (GP-IB)
Header: ON, separator: ";", terminator: "LFEOI", the comparator table
number set by "CSET" commands: 1, the temperature conversion table
number set by "DSET" commands: 1
Items which are not affected:
Status byte register, standard event status register, each enable register,
interface function, GP-IB address, output queue, input buffer, and current
path
*
RST

Table of Contents

Related product manuals