8.4 Command Reference
────────────────────────────────────────────────────
Synta
・ Executes the zero adjustment, and returns its result as 0 or 1 (data).
・ data = 0 means that the zero adjustment has been completed.
・ data = 1 means that the measurement value of the wiring resistance has
exceeded 100 count.
Respons
If headers are ON
If headers are OFF
:ADJUST data
data
Erro
The execution error is generated in the following circumstances.
・ When the measurement mode is not set to the resistance measurement.
・ When the resistance measurement range is set to the auto range.
・ Under the holding state.
Exampl
Transmission
Response
:ADJ?
:ADJUST 0
:ADJ?
0
The zero adjustment has been executed, and completed normally.
Synta
0 to 15 (numerical data in NR1 format)
Functio
Sets the comparator table intended to use as follows.
・ When "data = 0", the comparator becomes OFF.
・ When "data = 1 to 15", the tables which are set to the comparator is used.
Erro
The execution error is generated in the following circumstances.
・ When the value out of the setting range, or the unset tables are selected.
・ When the measurement mode is set to the temperature measurement.
・ When using the auto range.
・ When displaying the temperature conversion, if one of tables 2 to 15 is
specified
Exampl
Transmission
:COMP 1
The comparator table of number 1 is used.
:ADJust?
:COMParator