120
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Queries the setting of the comparator table for channel 1.
Syntax
COMParator1:SET?
(COMParator:SET? is also available)
Function
Returns the current setting of the comparator table for channel 1.
The table number is current setting number.
The channel number need not be specified, as channel 1 is presumed by
default.
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":COMPARATOR1:SET_<ON/OFF>"
"<ON/OFF>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":COMP1:SET?"
"COMPARATOR1:SET ON"
"ON"
Sets the table number for comparator measurement on channel 1 of the comparator.
Syntax
COMParator1:TABLe_<NR1>
<NR1>
1 to 9 (numerical data in NR1 format)
Function
Sets the table number for comparator measurement on channel 1 of the
comparator.
The channel number need not be specified, as channel 1 is presumed by
default.
<NR1> can be received in NRf format, but the decimal fraction is rounded
(rounding up a final 5).
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not in NRf format.
Device dependent error/
In the event of a system error
If this command is executed in the hold state (HOLD indicator lit or
flashing)
Example
Transmission
":COMP1:TABL 1"
Sets the table number to 1.
COMParator1:SET?
(Short form: COMP1:SET?)
COMParator1:TABLe
(Short form: COMP1:TABL)