8.4 Command Reference
────────────────────────────────────────────────────
Queries the setting of table number for comparator measurement on channel1of
the comparator.
Syntax
COMParator1:TABLe?
Function
Returns the current setting of the table number for comparator measurement on
channel 1 of the comparator.
The channel number need not be specified, as channel 1 is presumed by
default.
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":COMPARATOR1:TABLE_<1-9>"
"<1-9>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":COMP1:TABL?"
":COMPARATOR1:TABLE 1"
"1"
Enables or disables the comparator function for channel 2.
Syntax
COMParator2_<ON/OFF>
Function
Enables or disables the comparator function for channel 2.
The channel number must be specified, as otherwise channel 1 is presumed by
default.
When the comparator for channel 2 is set to all OFF, the comparator table
when this command is received is enabled.
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not character data.
Device dependent error/ In the event of a system error
Example
Transmission
":COMP2 ON"
Enables the comparator function for channel 2.
COMParator1:TABLe?
(Short form: COMP1:TABL?)
COMParator2
(Short form: COMP2)