8.4 Command Reference
────────────────────────────────────────────────────
Sets event status enable register 1. (Event for measurement value)
(Refer to Section 8.3.13.)
Syntax
ESE2_<NR1>
<NR1>
0to255
Function
Sets event status enable register 2 (ESER2) to the bitmask for controlling
access to events in event status register 2 (ESR2).
The decimal fraction is rounded (rounding up a final 5).
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
CODI CLO2 CIN2 CHI2 CLO1 CIN1 CHI1
Event Status Enable Register 2 (ESER2)
Note
Directly after the power has been switched on, or after a reset caused as a
result of key operation, the value is initialized to 0.
Bit7 is disregarded.
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not in NRf format
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
":ESE2 5"
Bits 0 and 2 of ESER2 are set to 1.
Reads out event status enable register 2. (Event for measurement value)
(Refer to Section 8.3.13.)
Syntax
ESE2?
Function
Returns the value of event status enable register 2 as a numerical value in NR1
format (0 to 127).
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":ESE2_<0-127>"
"<0-127>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":ESE2?"
":ESE2 5"
"5"
ESE2
ESE2?