138
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
■ Sets event status enable register 1. (Event for measurement value)
(Refer to Section 8.3.13.)
Syntax
ESE2_<NR1>
<NR1>
0to255
Function
・ Sets event status enable register 2 (ESER2) to the bitmask for controlling
access to events in event status register 2 (ESR2).
・ The decimal fraction is rounded (rounding up a final 5).
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
CODI CLO2 CIN2 CHI2 CLO1 CIN1 CHI1
Event Status Enable Register 2 (ESER2)
Note
・
Directly after the power has been switched on, or after a reset caused as a
result of key operation, the value is initialized to 0.
・
Bit7 is disregarded.
Errors
・
Execution error/ If the setting data is other than listed above.
・
Command error/ If the setting data is not in NRf format
・ System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
":ESE2 5"
Bits 0 and 2 of ESER2 are set to 1.
■ Reads out event status enable register 2. (Event for measurement value)
(Refer to Section 8.3.13.)
Syntax
ESE2?
Function
Returns the value of event status enable register 2 as a numerical value in NR1
format (0 to 127).
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":ESE2_<0-127>"
"<0-127>"
Errors
・ Query error/ If the response message is longer than 1000 bytes.
・ System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":ESE2?"
":ESE2 5"
"5"
ESE2
ESE2?