139
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
■ Reads out event status register 0. (Event for unit settings and power integration
value) (Refer to Section 8.3.13.)
Syntax
ESR0?
Function
Returns the value of event status register 0 (ESR0) as a numerical value in
NR1 format, 0 to 255, and then clears event status register 0.
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
DS FOR OT IE COR IDO PODI MODI
Event Status Register 0 (ESR0)
Notes
・ No header is prefixed to the response message.
・ With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON/OFF
"<0-255>"
Errors
・ Query error/ If the response message is longer than 1000 bytes.
・ System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
":ESR0?"
"16"
The integrating end (IE) flag is raised.
ESR0?