140
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
■ Reads out event status register 1. (Event for measurement value)
(Refer to Section 8.3.13.)
Syntax
ESR1?
Function
Returns the value of event status register 1 (ESR1) as a numerical value in
NR1 format, and then clears event status register 1. The numerical value is
one of the set: 0 through 255.
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
AOW AOA AOV OA OV HW HA HV
Event Status Register 1 (ESR1)
Note
・ No header is prefixed to the response message.
・ With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON/OFF
"<0-255>"
Errors
・ Query error/ If the response message is longer than 1000 bytes.
・ System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
":ESR1?"
"24"
The peak voltage out of range flag (OV) and peak current out of range flag
(OA) are raised.
ESR1?