152
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Sets the rectifier type.
Syntax
RECTifier_<NR1>
<NR1>
1, 2, 3 (1: RMS, 2: MEAN, 3:MEAN+FILTER)
Function
Sets the rectifier type.
<NR1> can be accepted in NRf format, but the decimal fraction is rounded
(rounding up a final 5).
Note
After changing the rectifier type, wait at least 5 seconds for the internal circuit
to stabilize before taking readings.
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not in NRf format.
Device dependent error/
In the event of a system error
If this command is executed in the hold state
(HOLD indicator lit or flashing)
During integrating (RUN indicator lit)
Example
Transmission
":RECT 1"
Sets the rectifier type to RMS.
Queries the rectifier.
Syntax
RECTifier?
Function
Returns the presently set rectifier type as a numerical value in NR1 format
(1 to 3).
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":RECTIFIER_<1/2/3>"
"<1/2/3>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":RECT?"
":RECTIFIER 1"
"1"
RECTifier
RECTifier?

Table of Contents