167
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Queries the settings of voltage range.
Syntax
VOLTage:SELect?
Function
Returns the settings of voltage range as a numerical value in NR1 format.
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":VOLTAGE:SELECT_<NR1>"
"<NR1>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":VOLT:SEL?"
":VOLT:SELECT 60"
"60"
Changes the power computation circuit clock.
Syntax
WCLock_<NR1>
<NR1>
0, 1, 2 (0: 2.37 kHz, 1: 2.44 kHz, 2: 2.604 kHz)
Function
Sets the power computation circuit clock.
<NR1> can be accepted in NRf format, but its effective value will be obtained
by rounding off the fourth digit on the basis of 5 and above being rounded up
and 4 and below being rounded down.
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not in NRf format
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
":WCL 0"
Sets the clock to 2.37 kHz.
VOLTage:SELect?
WCLock

Table of Contents