EasyManua.ls Logo

Hioki 3332 - Page 182

Hioki 3332
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
168
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Queries the power computation circuit clock.
Syntax
WCLock?
Function
The current setting of power computation circuit clock. is returned as a
numerical value in NR1 format.
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":WCLOCK_<0/1/2>"
"<0/1/2>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":WCL?"
":WCLOCK 0"
"0"
WCLock?