206
────────────────────────────────────────────────────
11.2 Standard Specifications
────────────────────────────────────────────────────
Analog output
Output voltage 5 VDC f.s. (U, I, P output simultaneously)
Output accuracy measurement accuracy ±0.2%f.s. (45 Hz or less, when
response time is SLOW)
Response time (suddenly change to 0% to 90% and 100% to 10% for
range)
FAST 0.2 to 0.3 seconds max. for voltage, current, active
power
SLOW 5 to 15 seconds for voltage, current, active power
Output resistance 100 Ω±5%
Monitor output
Output voltage 1 VACf.s. (u, i output simultaneously)
Output accuracy measurement accuracy ±0.1%f.s. (rms value level)
Output resistance 100 Ω±5%
Temperature coefficient ±0.02% f.s./℃ max.
External magnetic field
influence *
±1.5% f.s. max. (in a magnetic field of 400 Arms/m, 50/60 Hz.)
Power factor influence ±0.4% rdg. max. (45 Hz to 66 Hz, power factor=0.5)
±0.23% f.s. max. (45 Hz to 66 Hz, power factor=0)
Crest factor
Voltage lower value of either (measurement range×6)/measurement
value or 1100 V/measurement value
Current lower value of either (measurement range×6)/measurement
value or 90 A/measurement value
Influence of maximum rated
voltage to earth *
±0.05%f.s. max. (short-circuited, and apply 600 Vrms, 50/60 Hz between
input terminal and earth)
Frequency characteristics
(Measurement accuracy) *
±0.1% rdg.±1 dgt. (0℃ to 40℃,32
。
F to 104
。
F), in sinewave input)
Characteristic of integrator
(Measurement accuracy)
Integrated value measurement accuracy of current and active power
±1 dgt.
Integration time ±100 ppm±1s(0℃ to 40℃,32
。
F to 104
。
F)
*: at 200 mA range or more
Table 1 Range Configuration Table
Range for each channel unit:W
I
U
1.0000 mA 2.0000 mA 5.0000 mA 10.000 mA 20.000 mA 50.000 mA 100.00 mA 200.00 mA
15.000 V 15.000 m 30.000 m 75.000 m 150.00 m 300.00 m 750.00 m 1.5000 3.0000
30.000 V 30.000 m 60.000 m 150.00 m 300.00 m 600.00 m 1.5000 3.0000 6.0000
60.000 V 60.000 m 120.00 m 300.00 m 600.00 m 1.2000 3.0000 6.0000 12.000
150.00 V 150.00 m 300.00 m 750.00 m 1.5000 3.0000 7.5000 15.000 30.000
300.00 V 300.00 m 600.00 m 1.5000 3.0000 6.0000 15.000 30.000 60.000
600.00 V 600.00 m 1.2000 3.0000 6.0000 12.000 30.000 60.000 120.00
unit:W
I
U
500.00 mA 1.0000 A 2.0000 A 5.0000 A 10.000 A 20.000 A 50.000 A
15.000 V 7.5000 15.000 30.000 75.000 150.00 300.00 750.00
30.000 V 15.000 30.000 60.000 150.00 300.00 600.00 1.5000 k
60.000 V 30.000 60.000 120.00 300.00 600.00 1.2000 k 3.0000 k
150.00 V 75.000 150.00 300.00 750.00 1.5000 k 3.0000 k 7.5000 k
300.00 V 150.00 300.00 600.00 1.5000 k 3.0000 k 6.0000 k 15.000 k
600.00 V 300.00 600.00 1.2000 k 3.0000 k 6.0000 k 12.000 k 30.000 k
For a
arent
ower
reactive
ower
re
lace the unit W with VA or var.