EasyManua.ls Logo

Hioki 3332 - Page 220

Hioki 3332
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
206
────────────────────────────────────────────────────
11.2 Standard Specifications
────────────────────────────────────────────────────
Analog output
Output voltage 5 VDC f.s. (U, I, P output simultaneously)
Output accuracy measurement accuracy ±0.2%f.s. (45 Hz or less, when
response time is SLOW)
Response time (suddenly change to 0% to 90% and 100% to 10% for
range)
FAST 0.2 to 0.3 seconds max. for voltage, current, active
power
SLOW 5 to 15 seconds for voltage, current, active power
Output resistance 100 Ω±5%
Monitor output
Output voltage 1 VACf.s. (u, i output simultaneously)
Output accuracy measurement accuracy ±0.1%f.s. (rms value level)
Output resistance 100 Ω±5%
Temperature coefficient ±0.02% f.s./ max.
External magnetic field
influence *
±1.5% f.s. max. (in a magnetic field of 400 Arms/m, 50/60 Hz.)
Power factor influence ±0.4% rdg. max. (45 Hz to 66 Hz, power factor=0.5)
±0.23% f.s. max. (45 Hz to 66 Hz, power factor=0)
Crest factor
Voltage lower value of either (measurement range×6)/measurement
value or 1100 V/measurement value
Current lower value of either (measurement range×6)/measurement
value or 90 A/measurement value
Influence of maximum rated
voltage to earth *
±0.05%f.s. max. (short-circuited, and apply 600 Vrms, 50/60 Hz between
input terminal and earth)
Frequency characteristics
(Measurement accuracy) *
±0.1% rdg.±1 dgt. (0 to 40,32
F to 104
F), in sinewave input)
Characteristic of integrator
(Measurement accuracy)
Integrated value measurement accuracy of current and active power
±1 dgt.
Integration time ±100 ppm±1s(0 to 40,32
F to 104
F)
*: at 200 mA range or more
Table 1 Range Configuration Table
Range for each channel unit:W
I
U
1.0000 mA 2.0000 mA 5.0000 mA 10.000 mA 20.000 mA 50.000 mA 100.00 mA 200.00 mA
15.000 V 15.000 m 30.000 m 75.000 m 150.00 m 300.00 m 750.00 m 1.5000 3.0000
30.000 V 30.000 m 60.000 m 150.00 m 300.00 m 600.00 m 1.5000 3.0000 6.0000
60.000 V 60.000 m 120.00 m 300.00 m 600.00 m 1.2000 3.0000 6.0000 12.000
150.00 V 150.00 m 300.00 m 750.00 m 1.5000 3.0000 7.5000 15.000 30.000
300.00 V 300.00 m 600.00 m 1.5000 3.0000 6.0000 15.000 30.000 60.000
600.00 V 600.00 m 1.2000 3.0000 6.0000 12.000 30.000 60.000 120.00
unit:W
I
U
500.00 mA 1.0000 A 2.0000 A 5.0000 A 10.000 A 20.000 A 50.000 A
15.000 V 7.5000 15.000 30.000 75.000 150.00 300.00 750.00
30.000 V 15.000 30.000 60.000 150.00 300.00 600.00 1.5000 k
60.000 V 30.000 60.000 120.00 300.00 600.00 1.2000 k 3.0000 k
150.00 V 75.000 150.00 300.00 750.00 1.5000 k 3.0000 k 7.5000 k
300.00 V 150.00 300.00 600.00 1.5000 k 3.0000 k 6.0000 k 15.000 k
600.00 V 300.00 600.00 1.2000 k 3.0000 k 6.0000 k 12.000 k 30.000 k
For a
pp
arent
p
ower
,
reactive
p
ower
,
re
p
lace the unit W with VA or var.

Table of Contents