209
────────────────────────────────────────────────────
11.3 Function Specifications
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
(5) Integrator (Indicated as Ah or Wh, TIME (integration time) on the display)
Integration item
I, ±P,+P,-P integrated simultaneously
The RUN lamp is lit during integration, flashing during stop,
turned off when integration reset
Number of integration
5 times/second
Integration time
10 seconds to 10000 hours (10 seconds step)
Display limit
0.00000 m to ±999999 M
Effective input range
Same as the effective input limits of current and active power
Error warning
When an "o.r" is included in the integration value, the unit (Wh,
Ah) will flash.
Synchronizing with other functions
Interval printing to the printer, start/stop and data reset of
integration by outside control, and saving the integration value by
backup batteries.
Power failure recovery
If there is a power failure during integration, the unit will resume
integration after the power is restored.
(6) Frequency Measurement (Indicated as Hz on the display)
Measurement item
Select one of U, I
Measurement range
500 Hz/100 kHz/auto-range
Display limit
500 Hz range: 0.2% to 100% (1.0000 to 500.00 Hz)
100 kHz range: 0.4% to 100% (400.00 to 100.00 kHz)
Out of range display
If out of display range, "o.r", out of range will be displayed.
Measurement Frequency
1 to 5 times/seconds
(7) Rectifier type changing (only voltage, the true r.m.s. value is always displayed for current.)
RMS
True r.m.s. value
MEAN
Average rectifier effectiveness value indicator
MEAN+FILT
Average rectifier effectiveness value indicator (filter 500 Hz)
(accuracy assurance: up to 60 Hz)