5.1 Electrical Power Integration and Printer Output Examples
────────────────────────────────────────────────────
c
d
Start integration
Integration value from the
integration start.
Integration value from the
previous to this time print output.
Stop integration
INTEGRATOR START
00000:00:00
INTEGRATOR
TOTAL TIME 00000:00:00
0.0000kWh(+)
INTERVAL TIME 00000:00:00
0.0000kWh(+)
00000:10:00
INTEGRATOR
TOTAL TIME 00000:10:00
0.6691kWh(+)
INTERVAL TIME 00000:10:00
0.6691kWh(+)
.
.
.
00001:00:00
INTEGRATOR
TOTAL TIME 00001:00:00
4.0157kWh(+)
INTERVAL TIME 00000:10:00
0.6694kWh(+)
INTEGRATOR STOP
Using the RS-232C interface
Setting example: baud rate: 9600, data length: 8, parity: no, stop bit: 1
1. Using the key, move the flashing point to display c.
2. Set the baud rate to "9600" on display c, using the ▲ or ▼ key.
3. Using the key, move the flashing point to display d.
4. Set the data length to "8" on display d, using the ▲ or ▼ key.
5. Press the key once to move the flashing point to the right.
6. Set the parity to "n", using the ▲ or ▼ key.
7. Press the key once to move the flashing point to the right.
8. Set the stopbit to "1", using the ▲ or ▼ key.
9. Press the key once to move the flashing point to the right.
10. Set the parity to "Pr", using the ▲ or ▼ key.
11. Press the
ENTER
key and the setting will be decided.
(8) Integration start
1. Press the
INTERFACE
or
EXIT
key to return the measurement display.
2. Press the
START/STOP
key.
3. The RUN indicator is lit and the integration starts. "INTEGRATOR START"
is printed on the printer. Integration will print the measurement value every
ten minutes, and will stop 1 hour later. (RUN indicator will flash, and
"INTEGRATOR STOP" will be printed by the printer)