5.2 PT and CT Setting (Examples of Changing the Digits and Moving the Decimal Point)
────────────────────────────────────────────────────
Voltage 9.0000 kV 150 V×60
Current 2.0000 kA 5A×400
Active power 18.000 MW (150 V×5A×60×400)
Apparent power 18.000 MVA (150 V × 5A× 60 × 400)
Reactive power 18.000 Mvar (150 V × 5A× 60 × 400)
Current integration 0.00000 kAh (On starting integration)
Power integration 0.0000 MWh (On starting integration)
(5) Move onto the measurement display.
1. Press the
PT・CT・SC
or
EXIT
key, and the SHIFT indicator will turn off. The
SC indicator (scaling) is lit.
2. The range setting corresponds to the secondary side of the voltage and
current transformers. For example, with the 150 V, 5 A ranges, the ranges
(resolution) of the scaled voltage, current, active power, and other values is
as follows.