8.6 Message Reference
Reading the Most Recent Measurement
Syntax
Query
:FETCh?
Description
Reads the most recent measurement. No trigger occurs.
Example
Query
Response
:FETC?
17.0216E-3
Note
The following HIOKI 3227 command can be used, but the format of the response
message is different.
:MEASure:RESIstance?
However, the long-form :MEASURE:RESISTANCE? command operates the same
as the :MEASure:RESistance? command of this model.
❖
Measure in a Specifying Range and Function (
Ω
, LP
Ω
) (Page 146)
Reading the Temperature Measurement
Syntax
Query
:MEASure:TEMPerature?
Description
Reads the most recently measured temperature value.
The temperature measurement can be read regardless of the current resistance
function.
Example
Query
Response
:MEAS:TEMP?
25.1.0000E+0
Measuring (Awaiting Triggers and Reading Measurements)
Syntax
Query
:READ?
Description
Switches from the Idle State to the Trigger Wait State, then reads the next
measured value. With auto-ranging enabled, the most suitable range is selected
before measurement.
Error
• This command causes an execution error if issued during the Continuous
Measurement state (after :INITIATE:CONTINUOUS ON).
• This command causes an execution error if issued during the Trigger Wait
State.
Note
• The next command does not execute until measurement is finished.
• With external triggering using the
∗TRG command, after sending the ∗TRG
command and waiting for a time equivalent to the sampling rate, specify the
Talker. (only with the GP-IB interface setting)
Trigger Source Operation
IMMediate Triggers and reads measured value.
EXTernal
After triggering by the TRIG terminal (External I/O),
∗TRG command
or TRIG key, reads the measured value.