10.2 Error Display
168
Cleaning _____________________________________________________
To clean the instrument, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.
10.2 Error Display
Display Description
Err02
Zero-Adjust Range Error The value before zero-adjustment exceeded 1,000 dgt.
Err10
Execution Error The data portion of a remote command is invalid.
Err11
Command Error The command portion of a remote command is invalid.
Err80
Manual Adjustment Range Error
The valid adjustment range was exceeded during
adjustment.
Err90
ROM Error An internal program error occurred. Repair is required.
Err91
RAM Error An internal RAM error occurred. Repair is required.
Err92
EEPROM (Adjustment Data) Error Adjustment data is corrupted. Repair is required.
Err95
Resistance A/D Communications
Error
The A/D converter used for resistance measurement is
damaged. Repair is required.
Err96
Temperature A/D Communications
Error
The A/D converter used for temperature measurement
is damaged. Repair is required.
ErrCur
Constant-Current Fault
ErrHi
SENSE-H Open Circuit
ErrLo
SENSE-L Open Circuit
- - - - -
Constant-Current Fault, SENSE-H Open Circuit and SENSE-L Open Circuit are occurring
simultaneously.