EasyManua.ls Logo

Hioki 3541 - 5.13 Panel Load Function

Hioki 3541
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.8 Self-Calibration
66
Function
Description
To enhance measurement precision, this instrument performs self-
calibration to compensate for internal circuit offset voltage and gain
drift.
With SLOW1 and SLOW2 sampling, self-calibration is performed once
for each measurement. The settings here do not apply when SLOW1
or SLOW2 is selected.
With FAST and MEDIUM sampling, to increase measurement speed,
self-calibration is performed only with the timing specified as follows.
Self-calibration: Auto
Self-calibration is performed for about 55 ms once every 30
minutes.
Self-calibration: Manual
Self-calibration is performed when the CAL terminal of the EXT I/O
connector is connected to GND.
Self-calibration should always be performed after warm-up.
5.8 Self-Calibration
Setting Auto or Manual Self-Calibration (FAST or MEDIUM)
1
(SHIFT Lamp lit)
The Menu display appears.
2
The self-calibration setting display appears.
(Refer to the Menu display (Page 15))
Select Auto or Manual on the Sub Display.
AUto..... Auto self-calibration
In .......... Manual self-calibration
3
Applies setting and returns to the Measurement display.
(Main Display)
(Sub Display)
Self-calibration
is performed in the following cases regardless of the
above settings:
When the range is changed
When the sampling rate is changed
When a Load operation (refer to 5.13 Panel Load Function (Page
72)) is executed
When a reset is performed
When the measurement function is changed.
To obtain the specified accuracy, perform self-calibration in the
following cases:
•After warm-up
When the ambient temperature changes by 2
°C or more
Triggers occurring during self-calibration are delayed so that the
corresponding measurement occurs only after self-calibration has
finished. When using an external trigger, measurement may occur at
unintended times, in which case we suggest selecting Manual self-
calibration.

Table of Contents

Related product manuals