9.2 Accuracy
166
Sampling _____________________________________________________
Resistance and Low-Power Resistance Measurement
During
measurement
(Trigger to EOC=ON)
• DELAY = 0 ms, OVC = OFF, TC = OFF, Statistical Calculation = OFF,
Comparator = Hi/Lo
• With FAST and MEDIUM sampling settings, AUTO self-calibration
(if enabled) occurs for 55 ±10 ms every 30 minutes.
During importing (from INDEX=OFF to INDEX=ON)
Temperature
Measurement
Sampling Rate: 400 ±10 ms
[ms]
Line Frequency SLOW2 SLOW1 MEDIUM FAST
50Hz 455±10 155±5 21±1 0.60±0.3
60 Hz 449±10 149±5 17±1
0.60±0.3
[ms]
Line Frequency SLOW2 SLOW1 MEDIUM FAST
50Hz 400±10 100±5 20.0±1 0.30±0.1
60 Hz 400±10 100±5 16.7±1 0.30±0.1