5.4 Temperature Correction Function (TC)
Enabling/Disabling Temperature Correction
TC lit .........Temperature Correction enabled
TC off ........Temperature Correction disabled
An error appears when
you press the
TC/t key
The 9451 TEMPERATURE PROBE may not be connected, or may
be connected incorrectly. If Temperature Correction cannot be
enabled, check the connections of the temperature probe.
Temperature Correction does not work if the temperature probe is
allowed to touch the test object. Only the ambient temperature of
the immediate locale should be used.
Install the temperature probe and allow at least 60 minutes warm-
up before measurement. Unless the test object and temperature
probe used for temperature correction measurement have been
allowed to completely stabilize at ambient temperature, large
measurement errors may occur.
If the temperature probe is not inserted all the way into the TC
SENSOR jack on the rear of the instrument, large measurement
errors may occur.