28
Performing the Zero Adjustment
When the zero adjustment is not normally performed
When [0 ADJUST ERROR] appears, the proper adjustment is not performed.
Check the short-circuit method of the measurement probe and perform the zero adjustment with a proper
method so that the zero adjustment data values come within the range given in the table below.
Return
Impedance
measurement
R X
3 m
Ω
range -0.1000 m
Ω
to 0.1000 m
Ω
-1.5000 m
Ω
to 1.5000 m
Ω
10 m
Ω
range -0.3000 m
Ω
to 0.3000 m
Ω
-1.5000 m
Ω
to 1.5000 m
Ω
100 m
Ω
range -3.000 m
Ω
to 3.000 m
Ω
-1.500 m
Ω
to 1.500 m
Ω
Voltage measurement
-0.10000 V to 0.10000 V
Disabling zero adjustment
Select [OFF] on the zero adjustment screen.
(When [OFF] is selected, zero adjustment will be disabled. To enable, perform zero adjustment again.)
Selection
Conrm
(or)
Cancel