6.3 Saving Data
99
3
6
Chapter 6 Saving & Loading Data
Waveform data is saved in files named WAVEnnnn.MEM in folders named
with the date of recording (YY-MM-DD), created in the [HIOKI_LR8431]-
[DATA] folder (p.91).
To save a partial waveform, specify the time span to save beforehand (p.86).
Saving Waveform Data (with the SAVE Key)
Set the [Priority media].
(CF card/USB flash drive)
Quick SaveSelect & Save
(Before pressing the SAVE key, make
settings on the System screen)
(Select from the dialog that appears when
you press the SAVE key)
Select [Save] and press the ENTER key.
Select [Yes] in the confirmation
dialog, and press the ENTER key.
Select [Waveform], and
press the
ENTER key.
Select [Waveform]
as the [Save Type].
( default setting)
For the [Format] setting, select the data saving format.
Binary Select this format to be able to reload the saved data into the instrument,
or so that it can be loaded into the Logger Utility program.
Text Select this format to save the data as a text file.
From the [Span] setting options, select the portion of the data to save.
All *, A-B, Start-A, Start-B, A-End, B-End
(All = all acquired waveforms, A = Cursor A, B = Cursor B,
Start = start of acquired waveforms, End = end of acquired waveforms)
1
2
12
1
2
Set the [Save media].
(CF card/USB flash drive)