8.3 Making System Settings
126
Three USB modes are available: [USB Memory], [USB Communication], and
[USB Drive].
Switching the USB mode
Before changing the USB mode, disconnect both the USB com-
munication cable and the USB flash drive from the instrumen
t.
Failure to do so will cause a message warning you to eject the
devices to appear on the screen.
Select the [System]
Screen.
Select [USB Memory],
[USB Communication],
or [USB Drive] from the
pull-down menu.
Apply
A confirmation dialog appears.
1
2
3
Select [Yes].
Apply
[USB Memory]
[USB Communication]
[USB Drive]
USB method of use
Connector
used
[System] screen USB
mode setting
Use a USB flash drive. Type A USB Memory (Default)
Communicate with the LR8431-20
and initiate measurement using the
Logger Utility software from a com-
puter (using a USB cable).
Type B USB Communication
Read files on a CF card that is con-
nected to the LR8431-20 from a
computer (using a USB cable).
Type B USB Drive