1 Select a channel (CH 1 to CH 10) for the
[Channel] setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Voltage] for the [Input] (input type)
setting.
4 Select the appropriate measurement ampli-
tude range for the
[Range] setting.
Setting options:
100mV, 1V, 10V, 20V, 100V, 1-5V
( default setting)
1
2
3
4
: On
: Off
Instrumentation and Control Device Measurements
For instrumentation control signal measurements, select [1-5 V]. This range corre-
sponds to the 4-20 mA output standard for instrumentation control. This setting automat-
ically sets 1- and 5-volt lower and upper limits, respectively, in the 10 V range.
Measurement accuracy is ±10 mV ±5 mV. To measure 4-20 mA current, connect a 250
shunt resistor between + and – analog input terminals (p. 32).
Use the scaling function to scale measurements to any arbitrary unit value for display (p.
58).