Appendix 10 Frequently Asked Questions
A33
Appendix
How can I acquire data contin-
uously starting on 1/4/2012,
daily from 9:
00 to 17:00, for
one month?
Set as follows:
Measurement Criteria
[Interval]-[Cont]: Off, 8 hr 0 min 0 s
[Repeat]: On
Timer
[Start]: On, 12 Year, 4 Mon, 1 Day,
9 Hour 0 Min 0 Sec
[Stop]: On, 12 Year, 5 Mon, 1 Day, 9
Hour 0 Min 0 Sec
[Interval]: On, 1 Day, 0 Hour 0 Min 0
Sec
3.3, "Setting Configu-
ration 1 – Recording
Settings" (p.42)
4.4, "Starting and
Stopping Recording
by Timer" (p.76)
Is it possible to measure on 20
channels using two synchro-
nized HiLoggers?
Measurement start can be synchro-
nized using an external trigger.
Sa
mpling can
not be synchronized.
9.5, "Synchronous
Measurements with
Multiple Instruments"
(p.135)
Question Remedial Action Reference page
Measuring
Question Remedial Action Reference page
What should I do about zero
offset when the input is short-
ed?
Execute zero adjustment to com-
pensate for offset w
hen the in
puts
are shorted.
2.5, "Zero Adjust-
ment" (p.35)
Even though a signal is only in-
put on CH1, why do similar
waveforms appear on unused
cha
nnels?
An open-circuit input terminal can be
affected by signals on other chan-
nels. By closing the input terminal
ci
rcuit, normal measurement is pos-
sible. If this is impractical, disable
unused measu
rement channels, or
short the + and - terminals together.
How to display numerical data
values?
Press WAVE/DATA repeatedly to
cycle through the Waveform (only),
Waveform + Numerical Values, and
Numerical Values (only) displays.
Numerical values can be displayed
i
n
a large font on the enlarged nu-
merical display.
Numerical values at specific loca-
tions can be displayed using A/B
cursors.
1.4, "Screen Configu-
rations" (p.18)
"Displaying Cursor
Values" (p.84)
"Viewing Input Sig-
nals as Numerical
Values" (p.87)
What happens to pulse inte-
gration when there are more
than one mil
lion pulses?
The counter saturates at
1,073,741,823 counts. If you expect
to exceed this count, we recommend
measuring with the Integration Mode
(
[Count]) set to Instantaneous
(
[Inst]), and summing later with a
spreadsheet program.
"Integration (Count)