3.5 Display Settings
57
3
3
Chapter 3 Setting before Measurement
The vertical display range can be defined
by upper and lower limits.
Using this method, the voltage range does
no
t
need to be selected because the opti-
mum range is selected automatically for the
disp
la
y. However, upper and lower limit set-
tings are ignored when magnification and
ze
ro
position are enabled to define the dis-
play range.
As the upper and lower limits in the 1-5V
range
are fixed, they cannot be changed.
50%
100%
0%
50%
100%
50%
100%
0%
0%
100%
50%
0V
Zero Position: 0%
Display range
(Zoom: x1)
Measurement
range
Default Setting
Input type Zoom
Zero
position
Input type Zoom
Zero
position
Voltage
x1 50
Integration (Count)
Temperature (Tc)
x20 0
Revolution (Revolve)
x1 0
Specifying the Vertical Display Range by Upper and Lower Lim-
its (expansion/compression)
1 Select [Upper/Lower] for the [Type] setting
item.
2 Specify the [Upper] value for the top of the
display.
3 In the same way, specify the [Lower] value
for the bottom of the display.
The display range and scaling settings are in-
terlinked.
The range is not automatically optimized when
the upper and lower limits are set on the wave-
form screen.
1
2
3
Default Setting
Input type Range
Voltage
-50m to 50m
Temperature (Tc)
0 to 100
Integration (Count)
0 to 5000
Revolution (Revolve)
0 to 5000
After making the settings,
press [OK].